SENSe:PERiod:APERture? [MINimum|MAXimum]
The following table shows the relationship between the integration
time and the measurement resolution.
Integration
Time(PLC)
Resolution
DEFAULT
COMPATIBLE
0.02
0.0001 x Full-Scale
0.0001 x Full-Scale
0.1
0.00001 x Full-Scale
0.00001 x Full-Scale
1
0.000001 x Full-Scale
0.000003 x Full-Scale
10
0.0000001 x Full-Scale
0.000001 x Full-Scale
4.1.8
Sensor Selection for Temperature Measurements
The multimeter supports both thermocouple and RTD. You need to
configure the multimeter for the right sensor type, units and
transducer before conducting temperature measurements.
RTD
Definition
If the RTD is used, you have to confirm the RTD settings (incl. sensor,
units and transducer) first throught the front penal operations or the
remote control operation. For the sensor options, they are: PT100,
D100, F100, PT385, PT3916, user-defined RTD, NTCT and SPRTD. For
the unit options, they are: °C, °F and K. For the transducer options,
they are: 2W RTD and 4W RTD.
If you need to change the factors that are used to calculate the
temperature in RTD, choose user-defined RTD in which you are able to
change any factors as you wish. The default factors used are listed in
Table 4-3
:
65