164
Vorne Output Values
The decimal points shown above are not actually transmitted in the data stream, but their
position is implied. The displays are configured to show the decimal point in this position.
*Time Deviation is output in two formats in the same data stream: 33
±
s.ss and 34
±
sss.sss.
Output for the 33
±
s.ss format will be +9.bb when the measured value e9.99 (b = blank).
It will be -9.bb when the measured value is less than -9.99.
Table E.1 lists the output for the 34
±
sss.sss format, which observes the following conventions for
out-of-range values and leading blanks. Decimal points are implicit and do not appear in the data
stream.
Time Deviation Range
Format (b=blank)
Below -999.999
-bbb.bbb
-999.999 to -100.000
-sss.sss
-99.999 to -10.000
-bss.sss
-9.999 to -0.001
-bbs.sss
+0.000 to +9.999
+bbs.sss
+10.000 to +99.999
+bss.sss
+100.000 to +999.999
+sss.sss
Above +999.999
+bbb.bbb
Table E.1: Vorne 34 Output, Time Deviation Values
E.4
Sample 1133A Vorne Output
The string below illustrates the actual broadcast of the 1133A configured to Vorne protocol.
Vorne Output
Address
Meaning
Value
1100
11
Out-of-Lock Time
0
44144216
44
UTC/Local Time
UTC 14:42:16
22-00021
22
Frequency Deviation
-00.021 Hz
33-003
33
Time Deviation
-0.03 seconds
34-
0030
34
Time Deviation
-000.030 seconds
66144215
66
System Time
UTC 14:42:15
7759979
77
System Frequency
59.979 Hz
8807413
88
System Phase
74.13 degrees
8909591
89
System Magnitude
95.91 Vrms
55250
55
Day of Year
250