Chapter
3
26
ML-7 Series/ML-9001A
Chapter
3
Installation
■
Head Unit Air Purge Installation (ML-7064A Only)
If the following installation conditions cannot be met, always install air purge equipment for the Head
Unit.
Ambient temperature:
5
°
C to 35
°
C (non-condensing)
Temperature fluctuation:
Max.
±
10
°
C/h
Ambient Relative humidity: 40% to 80%
Oil mist, dust:
Max. 0.1 mg/m
3
•
Air purging makes the Head Unit less susceptible to the effects of dirt, dust, humidity, oil mist, and the
like by blowing clean air in the head and increasing the internal pressure. This helps prevent
problems with the optical system, such as soiling or scorching of mirrors.
•
When purging the air, supply air that meets the following conditions:
•
Recommended air purge unit (manufactured by SMC Corporation)
Mist separator:
AM150-02-BC
Heatless air dryer:
ID200-02B
Micro mist separator: AMD150-02-BC
Regulator:
AR2000-02BG
Hoses, cocks, and other required parts
Air volume
At least 0.1 m
3
/min.
Air pressure
29x10
4
to 88x10
4
Pa (3 to 9 kgf/cm
2
)
The recommended pressure is 39x10
4
Pa (4 kgf/cm
2
).
Do not supply air at a pressure exceeding 88x10
4
Pa (9 kgf/cm
2
). This may damage the
head.
Dew-point
temperature
Under -20
°
C
Use a regulator to adjust the pressure in the Head Unit so that it is slightly higher than
atmospheric pressure.
Head Unit
(Right-Angled Head)
Port for purging air
from the Head Unit
(RC 1/8 female)
Stop Valve
Regulator
Regulator
Mist Separator
Micro Mist
Separator
Heatless
Air Dryer
AC100V 50/60Hz
The user is
responsible
for providing
these items.
0.1 m
3
/min
29 x 10
4
Pa-
88 x 10
4
Pa
5 to 40°C