Programming Command Guidelines
4
N9360A Multi UE Tester W-CDMA Programming Manual
4-17
The Standard Event Status Register contains the following
information:
To query the Standard Event Status Register, send the
command
*ESR?
. The response is the decimal sum of the bits
which are set to
1
. For example, if the bit 7 and bit 3 are set to
1
,
the decimal sum of these 2 bits is 136 (128 plus 8) see
. So the decimal value 136 is returned.
Figure 4-7
Standard Event Status Register
Bit Number
D
es
cr
ip
ti
o
n
P
ow
er
O
n
U
nu
se
d
C
om
m
an
d
E
rr
or
E
xe
cu
tio
n
E
rr
or
D
ev
ic
e
D
ep
en
de
nt
E
rr
or
Q
ue
ry
E
rr
or
U
nu
se
d
O
pe
ra
tio
n
C
om
pl
et
e
7
6
5
4
3
2
1
0
*
ESR?
Table 4-5
Status Byte Register
Bit
Description
0
If all pending operations is completed following execution of the
*OPC
command, this bit is set to
1
.
1
This bit is always set to
0
.
2
If a query error has occurred, this bit is set to
1
.
3
If a device dependent error has occurred, this bit is set to 1. Device dependent errors have error numbers from
–399 to –300 and 1 to 32767.
4
If an execution error has occurred, this bit is set to
1
.
Execution errors have error numbers from –299 to –200.
5
If a command error has occurred, this bit is set to
1
.
Command errors have error numbers from –199 to –100.
6
This bit is always set to
0
.
7
If the instrument has been turned off and then on, this bit is set to
1
.
Summary of Contents for N9360A
Page 1: ...Agilent Technologies Agilent N9360A Multi UE Tester W CDMA Programming Manual...
Page 4: ...1 4 N9360A Multi UE Tester W CDMA Programming Manual 1 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 8: ...8 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 12: ...6 N9360A Multi UE Tester W CDMA Programming Manual THIS PAGE IS INTENTIONALLY LEFT BLANK...