Controlling the Sampler
4
Installation, Operation, and Maintenance
95
Part 2, Operation
Wash parameters
1
Display this screen.
Status / Automation / Injector / Washes
2
The syringe can be washed with sample before drawing up the amount to be
injected. It can also be washed with solvent both before (pre-washes) and
after (post-washes) an injection. The order of events is:
•
Wash syringe with solvent A
Solvent-A-Pre-Washes
times
•
Wash syringe with solvent B
Solvent-B-Pre-Washes
times
•
Rinse syringe with sample
Sample-Pre-Washes
times
•
Draw up sample and make injection
•
Wash syringe with solvent A
Solvent-A-Post-Washes
times
•
Wash syringe with solvent B
Solvent-B-Post-Washes
times
3
Enter your choices, then press
Esc
to return to the previous screen.
Needle depth (Sampling depth offset)
1
Display this screen.
Status / Automation / Injector / Depth Offset