22
Installation, Operation, and Maintenance
Part 1, Installation
1
Introduction to Installation
Overview
The controller
The 6850 and 6890N GCs contain an internal controller for the ALS.
The 6890 Plus GC has a slot that accepts a G2612A ALS Interface Board. This
adds controller capability to the GC.
The 5890 Series II and the 6890A GCs require an external G2912A controller.
The controller is programmed via a PC connection (5890 Series II) or the
setpoints are entered using the GC keypad (6890A).
The injector
The G2913A injector is designed for the ALS system. The G2613A injector may
also be used, but a firmware upgrade is required to use some of the features
described in this manual set.
The tray
Samples may be handled entirely with the 1-sample and 8-samples turrets
supplied with the G2913A injector. An optional G2916A (5890 Series II or
6890 using the G2917A upgrade kit) or G2614A sample tray increases the
capacity to 100 samples.
The Bar Code Reader
The optional G2615A Bar Code Reader (BCR) provides a positive sample
identification check. The BCR is not available for a 5890 Series II GC. Earlier
BCRs are not supported.