86
Installation, Operation, and Maintenance
Part 2, Operation
4
Controlling the Sampler
After setting required parameters, select
Send Settings to ALS
to load the values
into the G2912A Controller. For more information, see the online Help.
If the controller somehow becomes corrupted with bad parameters, it may be
necessary to reset the controller to its factory defaults. To restore the G2912
Controller factory defaults, turn the controller power off, press and hold the
[
Start
] and [
Stop
] buttons, turn the power on, hold the buttons in for five
seconds and release the buttons.
See
on page 131 for information on using the
[
Start
] and [
Stop
] buttons on the controller and 5890GC.
Figure 37
5890 ALS controller software main screen