46
Installation, Operation, and Maintenance
Part 1, Installation
2
Installation
Procedure 4. Connecting Cables
Figures
through
show the cabling required for installing a 7683B system
on a GC.
5890 Series II GC
Follow the instructions below to properly wire the G2912A controller. See
“Procedure 8. Install the G2912A ALS Controller”
on page 55 for installation
instructions for the G2912A controller including the power supply cord.
1
Connect the injector(s) to the controller.
2
Connect the tray to the controller with a G2614-60610 cable.
3
Connect the PC to the controller with a G1530-60600 cable.
4
Connect the controller to the GC with a G2912-60507 cable.
5
Optionally connect the 3396 Integrator to the 5890 GC with a 35900-60601
cable and to the controller with a 3396-61010 cable. See your 3396
Integrator documentation for wiring the analog signals and power supply.
6
Connect the power cords.
Figure 17
Cables for 5890 Series II GC
Power cords not shown
Injectors
G1530-60600
RS-232
5890 Series II GC
Remote
5890
G2912-60507
Tray
G2614-
G2912A Controller
60610
Tray
Front
Injector
Front
Back
Back
Injector
RS-232
PC with
Windows
Operating
System
Remote
6890
Remote
3396 Integrator
35900-60601
3396-61010
Summary of Contents for agilent 7683B
Page 10: ...10 Installation Operation and Maintenance...
Page 18: ...18 Installation Operation and Maintenance...
Page 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Page 72: ...72 Installation Operation and Maintenance...
Page 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Page 202: ...202 Installation Operation and Maintenance A Cables and Connectors...