Installation, Operation, and Maintenance
Part 2, Operation
5
Syringes and Needles
7
Move the plunger carrier loop down and tighten the plunger screw until it
stops.
8
Move the plunger carrier loop up and down. If the syringe plunger does not
move along with the carrier, repeat the previous steps. Be sure the plunger
screw is tight.
9
Check that the needle is aligned with the needle guide in the foot by moving
the slide up and down. The needle should slide smoothly in the needle
guide. See
C A U T I O N
Repeating this movement can damage the syringe.
Figure 42
Syringe carriage and needle support with syringe installed
Syringe latch (closed)
Needle support