Syringes and Needles
5
Installation, Operation, and Maintenance
105
Part 2, Operation
10
Pull down the syringe carriage until the needle support guide tip is near the
top of the inlet septum nut.
The needle support guide should be nearly centered over the tapered hole
in the septum retainer nut.
11
If the needle is not nearly centered over the septum retainer nut, check that
the syringe is installed correctly in the syringe carrier, the syringe needle is
straight, and the needle support assembly is properly installed.
C A U T I O N
Do not operate the injector without a syringe in place because the syringe latch may
interfere with the motor if it is allowed to swing freely.