Installation, Operation, and Maintenance
Part 1, Installation
2
Installation
Procedure 9. Replacing the Inlet Chassis Cover (6890A)
The original 6890A inlet chassis cover is not compatible with 7683B
components. (The inlet chassis cover is the blue plastic cover mounted over
the two inlet ports.) Replace the original cover with the one supplied with the
G2912A ALS Controller.
To remove the cover, remove the six Torx T-20 screws on the top of the cover
and lift it off. Install the new G1530-40070 cover in its place. See
Make sure all tubing is in the correct locations.
If you will be installing a tray, you may remove the three marked screws.
Figure 29
Installing the G1530-40070 inlet chassis cover
Remove if
installing
a tray