144
Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
8
Maintenance
Removing a Syringe
1
Loosen the plunger screw and raise the plunger carrier loop off the syringe
plunger.
2
Open the syringe latch.
3
Carefully lift the syringe flange out of the flange guide until clear, then lift
the syringe needle out of the needle support guide.
C A U T I O N
Be careful not to bend the syringe needle. Only pull the syringe out of the carriage until
clear. The needle bends easily when still seated in the needle support guide.