126
Installation, Operation, and Maintenance
Part 2, Operation
6
Vials and Bottles
Using a single solvent bottle of each type, all four waste bottles must be used;
41 vials can be run. Using two solvent A bottles and two solvent B bottles,
84 vials can be run. If you use three solvent A bottles and three solvent B
bottles, you can run 100 vials which is the tray capacity limit.
Table 19
For Solvent A and Waste A bottles
Total number of waste or
solvent bottles
Sample vials/Waste
bottles
Sample vials/Solvent vial
1
NA
41
2
100
84
3
NA
100
Table 20
For Solvent B and Waste B bottles
Total number of waste or
solvent bottles
Sample vials/Waste
bottles
Sample vials/Solvent vial
1
NA
62
2
100
100
3
NA
100
C A U T I O N
We strongly recommend that
all
waste positions in
all
turrets contain bottles at
all
times, regardless of which ones are actually used in the current configuration. This
protects against accidental discharge of solvent into the turret body, which could
damage it.