Installation
2
Installation, Operation, and Maintenance
65
Part 1, Installation
Procedure 11. Configuring Your GC and Data System
GC
When installation is complete, configure your GC for use with the 7683B
Automatic Liquid Sampler. See
details. Be sure to check:
•
Injector usage
•
Solvent bottle usage
•
Solvent volume usage
For 6890 GCs, be sure to configure the front injector as
Inj1
and the back
injector as
Inj2
.
Data system
Configuration
Agilent ChemStation and Cerity data systems contain information about the
sampling equipment in use. This must be updated to remove the old
information and replace it with information about the new equipment that you
have installed. See the data system documentation for details.
Note that for a 5890 GC system, neither ChemStation nor Cerity control the
sampler. Simply deconfigure the old injector and modify your methods to use
manual injection.
Upgrading methods
Before using methods created for a 7673 sampler, be sure to edit them as
needed to match the new hardware.