Faults and Errors
9
Installation, Operation, and Maintenance
161
Part 3, Maintenance and Troubleshooting
Fault light is on.
•
The injector door is open.
•
Turret is not recognized if door
is closed.
1
Ensure that the injector door is
closed.
2
Re-install turret.
3
If the fault light stays on, obtain
Agilent service.
Fault and Run lights
are on.
•
The injector is mounted
incorrectly on the mounting
post.
•
Incorrect mounting post.
1
Ensure that the injector is
mounted properly. For more
information, see
2
Ensure the correct mounting post
is installed. See
3
If the Fault and Run lights stay on,
obtain Agilent service.
Align Mode light is on.
•
The system was not initialized.
•
There is an injector memory
error.
•
Customer pushed the align
mode button.
1
Verify the turret is properly
installed. See
2
Perform the alignment procedure
to initialize the system. See
3
If alignment fails, obtain Agilent
service.
All lights are on
•
There is a board failure.
•
There is a firmware revision
conflict.
1
Check all cable connections.
2
Turn the instrument off, then on
again.
3
If the lights remain on, obtain
Agilent service.
Table 24
Status lights (continued)
Lights
Possible cause
Action
Summary of Contents for agilent 7683B
Page 10: ...10 Installation Operation and Maintenance...
Page 18: ...18 Installation Operation and Maintenance...
Page 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Page 72: ...72 Installation Operation and Maintenance...
Page 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Page 202: ...202 Installation Operation and Maintenance A Cables and Connectors...