110
Installation, Operation, and Maintenance
Part 2, Operation
6
Vials and Bottles
Preparing Sample Vials
Selecting sample vials
The 7683 Automatic Liquid Sampler injector and the tray use clear or amber
glass sample vials with crimp caps, or Target® DP™ screw-cap vials. Use
amber glass vials for light-sensitive samples. Refer to your Agilent catalog for
consumables and supplies for acceptable vial types. Incompatible sample vials
cause tray and turret errors.
shows the critical dimensions for sample vials and microvial inserts
to be used with the 7683 Automatic Liquid Sampler. These dimensions do not
make up a complete set of specifications.
Figure 44
Sample vial and microvial insert dimensions
5 minimum
3.7
8.2 ± 0.2
32 ± 0.5
11.7 ± 0.2
34.5 maximum
Microvial insert
Crimp cap sample vial
Maximum height of a capped vial
All dimensions in millimeters
6.25 ± 0.25