84
Installation, Operation, and Maintenance
Part 2, Operation
4
Controlling the Sampler
4
Load the syringe with sample.
a
Rotate the sample vial under the
syringe.
b
Lower the syringe needle so that it
pierces the vial septum.
c
Draw in sample.
d
With the needle still in the sample,
depress the syringe plunger quickly.
e
After the final pump, draw in
sample.
f
Raise the syringe needle out of the
sample vial.
•
Actions
c
and
d
can be repeated
several times. The purpose is to
expel air bubbles from the syringe.
5
Inject the sample.
a
Rotate the turret to expose the GC
inlet.
b
Lower the syringe needle so that it
pierces the inlet septum.
c
Depress the syringe plunger to
make the injection.
d
Raise the syringe needle out of the
GC inlet.
• Start
signals are sent to the GC and
the data processor at the moment
of injection.
6
Wash the syringe with solvent.
•
, but according to
the post-injection parameters.
7
Transport the sample vial.
•
Return the sample vial to its initial
turret or tray location.
8
Perform repeat injections.
•
If so programmed, wait for the GC to
become Ready and repeat the cycle
from
Table 13
Sampler cycle (continued)
Step
Action
Notes
Summary of Contents for agilent 7683B
Page 10: ...10 Installation Operation and Maintenance...
Page 18: ...18 Installation Operation and Maintenance...
Page 70: ...70 Installation Operation and Maintenance Part 1 Installation 2 Installation...
Page 72: ...72 Installation Operation and Maintenance...
Page 128: ...128 Installation Operation and Maintenance Part 2 Operation 6 Vials and Bottles...
Page 202: ...202 Installation Operation and Maintenance A Cables and Connectors...