174
Installation, Operation, and Maintenance
Part 3, Maintenance and Troubleshooting
10
Troubleshooting
Symptom: No Signal/No Peaks
Table 30
Signal/Peak problems
Possible cause
Action
Syringe plunger is malfunctioning.
Verify that the syringe plunger is secured by the plunger
screw. If the plunger screw is loose, tighten it. See
Check the syringe needle for plugging. If the syringe is
plugged, replace or clean the syringe.
Sample level is too low in vial.
If there is no or very little sample in the vial, the needle may
not be able to reach it. See
Alternately, you may edit your method to adjust the needle
sampling depth. See Sampling offset in
Sample is viscous.
If the sample is viscous, try the following:
•
Increase the viscosity delay time.
•
Use the tray quadrants to warm the sample.
•
Dilute the sample in an appropriate low-viscosity
solvent.
•
Turn the tower fan off.