Agilent 1290 Infinity LC System Manual and Quick Reference
4
System Setup and Installation
Installing the Module
Flow Connections Between Modules
When connecting the modules, always flush each capillary and the column
with solvent before connecting to the next component in the flow path.
1
Connect the outlet from the Jet Weaver mixer to the autosampler using a
0.12 mm i.d. stainless steel flexible capillary (color code label is red). This
should be connected to port #1 on the autosampler injection valve.
2
Connect a 0.12 mm i.d. stainless steel flexible capillary from port #6 on the
autosampler injection valve to the thermostatted column compartment.
Connect the capillary direct to the low dispersion volume heat exchanger
for minimal delay volume or to the switching valve if installed.
3
Connect the outlet of the column to the inlet (marked CELL-IN, left-hand
side connection) of the Max-Light Cartridge Cell in the 1290 Infinity
detector
4
Connect the waste tubing to the outlet (marked CELL-OUT, right-hand side
connection) of the Max-Light Cartridge Cell in the 1290 Infinity detector
and place the outlet in an appropriate waste collection container.
Integration Into the Network
For network integration of your system refer to user manuals of your modules
(chapter
LAN Configuration
).