Setting Up a Method using Edit Entire Method
Diode-Array Detector (DAD) tab
Figure 41
Setup Method screen – Diode-Array Detector tab
• Signals
: Up to eight separate signals (chromatograms) can be recorded. To
mark a signal for collection the
Use Signal
box is checked for that signal, the
wavelength and bandwidth are defined, and if a reference signal is required
that box is checked and defined.
•
Wavelength
sets the central wavelength (nm) of the signal,
•
Bandwidth
sets the width (nm) of the signal,
•
Reference Wavelength
sets the central wavelength (nm) of the reference
band which is subtracted from the analytical signal,
•
Reference Bandwidth
sets the width (nm) of the reference band.
• Peakwidth
sets the data collection rate and signal filtering.
• Stop Time
/
Post Time
are set to
No Limit / Off
and these values are usually
taken care of in the
Pump
tab. However, the detector module Stop Time can
sometimes differ from the pump Stop Time when it is required that data