Agilent 1290 Infinity LC System Manual and Quick Reference
3
Optimization of the Agilent 1290 Infinity LC System
How to Configure the Optimum Delay Volume
Replacing the Jet Weaver in the 1290 Infinity Binary Pump
1
Remove capillary connections from the Jet Weaver.
2
Remove the hex screws that fix the Jet Weaver to the
pump housing.
3
Install the new Jet Weaver.
N O T E
The Jet Weaver has a front and a rear side with
different internal volumes (35 / 100 µl) that are
optimized for a low delay volume or best mixing
performance.
4
Reinstall the capillary connections.
The inlet at the bottom of the Jet Weaver is connected to the
central port of the pump valve by a capillary (length 300 mm,
0.17 mm i.d.). The outlet at the top is connected to the
autosampler.