Agilent 1290 Infinity LC System Manual and Quick Reference
6
Appendix
Setting Up a Method using Edit Entire Method
The
Edit Integration Events
screen has two tables:
• Initial Events For All Signals
contains events (integration parameters) that
apply to all signals acquired with the method,
• Specific Events For Signal
contains events which are specific for one type of
detector or specific to different signals from the same detector.
Key parameters in this table are:
•
Slope Sensitivity
represents the slope and curvature of the baseline
needed to mark the start and end of a peak;
•
Peak Width
: the width at half-height of the narrowest peak of interest
should be entered. This helps the integrator to distinguish between noise
and very small peaks;
•
Area Reject / Height Reject
values that control rejection from the results of
peaks whose area or height falls below these values;
•
Integration OFF/ON
suppresses integration between set limits. Almost
always used to inhibit integration in the region from injection to solvent
front or unretained peak marker.
Lines such as
Integration OFF/ON
are added to the table using the icons at the
top of the window.
Click
OK
to exit and the next screen in the Edit Entire Method process will
open.
Specify Report
The
Specify Report
screen can also be directly accessed in
Method and Run Control
view by right-clicking on the graphical user interface on the Report icon and
then selecting
Specify Report
in the context menu. In the
Data Analysis
view it
can be accessed through the menu
Report > Specify Report
or the Specify Report
task icon.