What to do in the event of malfunctions
Surgical field illumination on
microscope not working.
Light guide not properly
inserted in microscope.
Insert light guide as far as it
will go.
Failure of main and backup
lamps.
Replace both lamps.
Stand electronics failure.
Illuminate surgical area with
an OR illuminator. Contact
your ZEISS Service.
Lamp module in stand has no
contact.
Insert lamp module to the
stop.
Light source not activated on
the foot control panel.
Press light source on/off
button on the foot control
panel.
Suspension arm is in the park
position.
Pull suspension arm
downward.
Insufficient surgical field
illumination.
Brightness level set too low.
Adjust brightness on
suspension system or foot
control panel.
Aging of the lamp can reduce
the amount of light.
Replace lamp
Defective light guide
(illumination not uniform).
Replace the light guide
Contact your ZEISS Service.
-
Light guide not correctly
inserted.
Check that the light guide is
connected.
Filter selection knob is in an
intermediate position.
Turn knob until filter is
correctly in place.
Lamp is not properly seated in
the lamp socket.
Properly push lamp into lamp
socket.
Surgical field illumination too
bright.
Brightness level set too high.
Adjust brightness on
suspension system or foot
control panel.
Switch off light source on
stand. Illuminate surgical area
with an OR illuminator.
Contact ZEISS Service.
Defective light control.
Switch to manual mode.
Fault
Possible reason
Corrective measure
Cross
reference
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...