G-30-1720-en
CAUTION
Never leave the device unattended when it is switched on!
If a patient or an object is exposed to the light source for a longer period of
time, this could cause burns or fire damage. Excessive radiation exposure
times may lead to retinal injury in the patient’s eye.
•
Never leave a device unattended when its light source is switched on.
NOTE
Restricted functions due to malfunctioning motor electronics!
If the motor electronics malfunction, major functions such as XY movement,
focus, zoom, and light control may be negatively affected.
•
Activate manual mode (see Page 131).
•
Move the microscope manually into the required position.
•
Set the magnification on the manual zoom adjustment knob.
•
Use an OR illuminator.
•
Contact ZEISS Service.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...