Version 8.0
G-30-1720-en
Page 15
OPMI Lumera i
Safety measures
During use
•
Only use the system after it has been properly balanced.
•
Constantly monitor the system during use.
•
Excessive radiation exposure times may lead to retinal injury in the patient’s
eye. Never leave the device unattended while its light source is switched on.
•
Avoid looking directly into the light source, e.g., into the microscope
objective lens or light guide.
•
When the illumination is switched on, there is a risk of fire or you may burn
yourself if the light guide is not connected at both ends.
•
Any type of radiation has a damaging effect on biological tissue. This is
also true for the light illumination the surgical field. You should therefore
reduce the brightness and the light exposure time to the absolute
minimum required.
•
Phototoxic effect of light beams. When operating on the eye, always use
the retina protection filter (blue barrier filter) installed in the light source
and/or the retina protection device to ensure that the patient’s eye is not
exposed to unnecessary (blue) radiation.
•
Adjust the illumination intensity and the resulting radiation exposure time by
selecting the appropriate illumination setting. The values recommended by
ZEISS are located in the table “Maximum radiation exposure times” on
Page 20. Deviations from these values must be medically justified and are the
sole responsibility of the surgeon.
•
Cables which are not appropriately routed may lead to an increased risk
of tripping during operation of the system. Always route cables so that
they do not hinder the user.
After every use
•
Use the main power switch to switch off the system.
•
The main power switch must always be off when the system is not in use.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...