Version 8.0
G-30-1720-en
Page 151
OPMI Lumera i
Device data
Ethernet connection (option), requirements for integrating the device into an
existing IT network
•
When connecting the device to an existing IT network, comply
with the following specifications and measures in accordance with
IEC 60601-1:
–
The device can be connected to an existing IT network in order to
transfer device parameters.
–
The IT network to which the device is to be connected must have the
following features:
–
IPv4, static or dynamic address
–
Any IP address.
–
The IT network to which the device is to be connected must be
configured as follows:
–
The firewall is open for inbound ports.
–
Data is not intended to be transmitted from the device to external
networks.
–
Requirements in relation to network connection with the device:
–
The IT network has an Ethernet connection via RJ45, at least Cat.5,
100 Mbit/s.
–
The device has a network isolator in compliance with IEC 60601-1
which means it is not connected to the network power supply.
–
The operator is responsible for the flow of information and the routing
between the device, the IT network, and other devices:
–
The essential performance features of the device do not require
a connection to the IT network or any data transfers.
–
The operator is responsible for the topology of the IT network and
the necessary routing.
If the IT network cannot provide the features required to integrate the
device into the IT network, the following hazards may arise:
–
Higher current at power output sockets; risk of electrical shock, if the user
touches a live part of the casing or a patient
–
Virus infestation of the system represents a hazard for the patient due to
possible prolongation of operating and anesthetic times
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...