Medical electrical
systems
Any additional equipment you connect to medical electrical devices must
demonstrably comply with the applicable standards and directives (e.g.,
IEC 60950-1 for data processing equipment).
In addition, all configurations must meet the standard requirements for
medical electrical systems (see IEC 60601-1).
Anyone connecting additional equipment to medical electrical devices is
a system configurer and as such is responsible for compliance of the system
with the normative requirements for medical electrical systems.
Please note that local legislation takes priority over the requirements of the
above-mentioned standards. If you have any questions, contact your local
dealer or ZEISS Service.
WARNING
Risk of serious injury due to unauthorized modification!
You must not modify this device without the permission of the manufacturer.
The manufacturer is not liable for damages caused by an unauthorized use
of the device.
Unauthorized modifications of the device may result in severe, irreversible
injuries to users and/or patients.
•
If you modify the device after consultation with the manufacturer, you
have to carry out the appropriate inspections and tests to ensure that it
can still be used safely.
R502
NOTE
Loss of warranty
The manufacturer
is not liable
for damages caused by an unauthorized use
of the device. Furthermore, such unauthorized modifications will render the
warranty and any claims under warranty invalid.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...