![background image](http://html1.mh-extra.com/html/zeiss/opmi-lumera-i/opmi-lumera-i_instructions-for-use-manual_4422443139.webp)
Version 8.0
G-30-1720-en
Page 139
OPMI Lumera i
Care and maintenance
Safety check
To prevent any impairment of the system’s safety as a result of aging, wear
etc., the operating company must ensure, in accordance with the applicable
national regulations, that the regular technical safety checks defined for this
device are performed on schedule and to the stipulated extent.
The safety checks must be performed only by the manufacturer or qualified
personnel.
At a minimum, the following safety checks should be performed on the system:
–
Availability of the Instructions for Use
–
Visual inspection of the device and accessories for damage, as well as
legibility of the labels
–
Leakage current test
–
Test of protective ground conductor
–
Function and wear test of the steerable casters and locking tabs
–
Function test of all switches, buttons, sockets, and indicator lamps of the
device
–
Function test of the Manual mode without the patient, at least semi-
annually
CAUTION
Risk of injury without safety inspection!
Hazards and device deficiencies will not be detected in time and can have
a negative effect on patients, users, or others.
•
Have a safety inspection in compliance with IEC 62353 performed on
time and to the prescribed extent. At the same time, be sure to comply
with the relevant national regulations.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...