![background image](http://html1.mh-extra.com/html/zeiss/opmi-lumera-i/opmi-lumera-i_instructions-for-use-manual_4422443136.webp)
What to do in the event of malfunctions
OPMI Lumera i
Version 8.0
Page 136
G-30-1720-en
Zoom error. The error can be
acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., jammed hardware
If the error impairs your work,
switch to the manual mode.
Focus error. The error can be
acknowledged.
If the error impairs your work,
switch to the manual mode
and contact ZEISS Service.
–
e.g., jammed motor of the
focus system
If the error impairs your work,
switch to the manual mode.
Error during system start.
Please reboot the system,
making sure that no function
buttons are pressed during
the booting process.
–
e.g., a button was pressed
during system start
Reboot the system. Make
sure not to press any buttons
during this process.
-
System error. The error can be
acknowledged.
If the error recurs, please
contact ZEISS Service.
–
Unknown software or
hardware error.
Acknowledge the error and
inform your ZEISS service
partner if it occurs again.
-
Error message
Possible reason
Corrective measure
Cross
reference
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...