Safety measures
OPMI Lumera i
Version 8.0
Page 26
G-30-1720-en
Locking lever for limiting vertical movement
4
Locking lever for limiting vertical movement
The locking lever is used to set the lower limit position of the suspension
arm. This prevents the surgical microscope from making contact with the
patient even if it is inadvertently lowered (see Page 45).
Position of the microscope with fundus imaging system
When using a wide-angle observation system (e.g., BIOM 3 from Oculus)
which is usually installed between the surgical microscope and the patient,
make sure that the patient is neither put at risk nor injured by the motorized
focusing system or the movement of the suspension system arm.
Only use accessories expressly certified by ZEISS for combination with the
surgical microscope described in this manual.
WARNING
Injury to the patient’s eye!
Incorrect operation of the fundus imaging system may cause injury to the
patient’s eye and impair image quality. Perform the following steps to avoid
this hazard:
•
Perform a focus reset.
•
With the fundus viewing system swung out of the beam path, always
position the microscope body in such a way that index dot (1) of the
microscope’s focus is in the middle of the upper triangle (2) of the marking.
•
Select a medium magnification (e.g., 1.0).
•
Lower the surgical microscope to the surgical field until the cornea of the
patient’s eye is in focus.
•
Turn the locking lever for limiting downward travel clockwise as far as it
will go; without any patients, check that the suspension arm cannot be
lowered any further.
•
It is vital that you read the Instructions for Use for the fundus imaging
system used (e.g., BIOM 3 from Oculus).
1
2
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...