Press the XY reset button on the X-Y coupling. The executed functions
depend on the configured settings (see Page 121).
•
Look through the eyepieces and lower the surgical microscope using the
suspension arm until the surgical field comes into focus. This will allow you
to achieve a rough focus.
•
Press the XY reset button on the X-Y coupling. The executed functions
depend on the configured settings (see Page 121).
•
Look through the eyepieces and lower the surgical microscope using the
suspension arm until the surgical field comes into focus. This will allow you
to achieve a rough focus.
CAUTION
Unintentional movement of the device
If a focus movement is triggered by incorrect operation, this can cause the
device to come into contact with the patient’s eye, especially if an accessory
(e.g., Visulux, BIOM, fundus viewing system) is installed on the surgical
microscope. This can result in injury to the anterior eye segment.
•
To prevent contact with the patient, ensure that the range of movement
is sufficient for the position of the focus.
WARNING
Risk of injury due to vertical movement of the suspension arm
If the suspension arm is adjusted, the suspension system can move vertically
causing the surgical microscope to come into contact with the patient.
This can result in serious injuries to the patient’s eye.
•
Ensure that there is adequate space for a vertical adjustment.
CAUTION
Video images not for diagnostic purposes!
R510
You must not use stored video sequences, video clips (extracted sequences),
single images, and live images for diagnostic purposes.
Since the video cameras and monitors have not been calibrated,
the visualized video sequences, video clips, and single images may include
deviations in scale, color, and shape. Transmission to other monitors, in
particular, may change the display of the images.
As a result, incorrect data would be used as a basis for subsequent treatment.
and lead to unnecessary or incorrect surgery.
•
Make sure that stored video sequences, video clips, single images,
and live images are never used for diagnostic purposes.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...