Preparation for use
OPMI Lumera i
Version 8.0
Page 64
G-30-1720-en
Configuring surgical microscope
Changing tubes, eyepieces, or objective lens
If a different tube, eyepiece, or objective lens is needed for an application, you
can change these components as required.
Changing tubes,
widefield eyepieces,
and the objective lens
•
Bring the suspension arm in a position convenient for you and tighten
friction adjustment knob (1).
CAUTION
Risk of injury caused by lowering of the surgical microscope!
•
Never prepare the system for use above the patient.
•
Never mount or remove accessory equipment on the surgical microscope
during surgery.
•
Only use accessories expressly certified by ZEISS for combination with the
surgical microscope described in this manual.
•
Never exceed the maximum permissible weight capacity of the suspension
arm (see Page 148).
•
Rebalance the suspension arm after every change of equipment (see
Page 73).
•
Tighten all friction knobs firmly.
WARNING
Risk of injury caused by falling parts!
•
Before every use and after re-equipping the system, make sure that
binocular tube (5) is securely locked in position. Make sure that securing
screw (4) and objective lens (3) or the objective lens carrier ring with
accessories have been firmly tightened!
CAUTION
Confusing the assistant tube with the main observer tube
If the assistant tube is incorrectly connected to the main observer interface,
the stereo image of the surgical field will not appear correctly, and the surgical
field will produce an incorrect stereoscopic impression.
If the surgery is performed with this configuration, this may cause injury to
the anterior segment of the eye.
•
Label the inverter tube as the assistant tube.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...