Safety measures
OPMI Lumera i
Version 8.0
Page 10
G-30-1720-en
•
Never look at the sun through the binocular tube, the objective lens, or an
eyepiece.
•
The maximum brightness of the lamp(s) can drop over time (normal device
attribute).
Ambient conditions
•
Do not use the power-operated devices included in the delivery package
–
in areas where there is a risk of explosion,
–
or at a distance of less than 25 cm from flammable anesthetics or
volatile solvents such as alcohol, benzine, or similar substances.
•
Do not use or store the device in a damp place. Do not expose the device
to dripping water, running water, sprayed water, or splashes.
•
Ensure that fluids cannot enter the device.
Electrical safety
•
Always switch off the device before you plug it into or unplug it from
a power socket, if you are not going to use it for a prolonged period of
time, or if you want to clean its surface.
•
Switch off the system at the power switch if you notice any smoke, sparks,
or unusual noise. Do not use the unit until it has been repaired by the ZEISS
Service team.
A potential equalization connector is provided on the instrument’s connection
panel (see IEC 60601-1). This connector can be used to connect other active
devices to the same ground potential or for redundant connection to the
protective ground.
•
Do not use multiple sockets!
•
Do not remove any parts of the housing! The system contains components
that carry high voltage and may cause electric shock.
•
Do not touch the device if your body is electrostatically charged and the
device is not grounded.
•
Before connecting the system to line power, make sure that the power
network is free from defects that may lead to dangerous voltage on the
power plug.
•
Do not pull on the light guide, power cord, or any other connecting
cables.
•
Please note the information on EMC (electromagnetic compatibility) in the
chapter "System data" on Page 158.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...