Operation
OPMI Lumera i
Version 8.0
Page 122
G-30-1720-en
•
Select the functions to be started when the XY reset button is pressed.
The following functions are available:
–
XY Reset
(The X-Y coupling moves to its center position.)
–
Focus Reset - always active
(The focusing system moves to its initial position.)
–
Zoom Reset
(The magnification is set to its preset start value.)
–
Light Reset
(The brightness is set to its preset start value.)
•
To save the changes, press the
button.
The setting is saved for the current user.
•
To return to the Main Menu, press the
button 2×.
The main menu is displayed again.
Configuring standby position
The standby position is reached when suspension arm (1) is moved to the
topmost position. This automatically deactivates the light source.
The functions are reset to their stored start values, depending on the settings
made. Proceed as follows for user-specific configuration.
•
Select <System Settings><Park Position> in the main menu.
The “Park Position” menu is displayed.
•
Select the functions to be performed when the standby position is reached.
The following functions are available:
–
XY Reset
(The X-Y coupling moves to its center position.)
–
Focus Reset - always active
(The focusing system moves to its initial position.)
–
Zoom Reset
(The magnification is set to its preset start value.)
–
Light Reset
(The brightness is set to its preset start value.)
1
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...