Care and maintenance
OPMI Lumera i
Version 8.0
Page 140
G-30-1720-en
Changing the halogen lamp
The following steps are required for changing the lamp:
•
Turn off the device at the power switch.
•
Press button (2) to eject lamp module (1) partially. Pull out the lamp module.
•
Remove the defective halogen lamp from the spring-loaded mount.
•
Pull the ceramic base (8) from the contact pins of the halogen lamp.
•
Plug the ceramic base (8) onto the contact pins of the new halogen lamp.
•
Insert new halogen lamp. Make sure you do not touch the lamp bulb (7)
or the interior of the reflector (6).
•
Press the halogen lamp into the spring-loaded mount (4)/(5).
•
Push the lamp module including the new halogen lamp back into the
device.
•
Turn on the device at the power switch.
3
“Instructions for Use” sign
Observe the Instructions for Use or further applicable documents.
9
“CAUTION” warning label
Warning of a potential hazard.
•
Follow the instructions to change the lamp.
NOTE
10 Risk of burns caused by hot lamp!
If you replace the lamp shortly after it has failed, the lamp will still be very hot.
•
Allow the lamp to cool down before removing it.
•
Always use heat-resistant protective gloves when replacing the lamp!
NOTE
Use the correct halogen lamp
•
Only use the 12 V, 100 W halogen lamp specified in “Ordering data,
Consumables”.
Содержание OPMI Lumera i
Страница 1: ...ZEISS OPMI Lumera i on the floor stand Instructions for Use G 30 1720 en Version 8 0 2018 11 26...
Страница 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Страница 25: ...Version 8 0 G 30 1720 en Page 25 OPMI Lumera i Safety measures Fig 2 Switch for manual mode 3 1 3...
Страница 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Страница 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Страница 37: ...Version 8 0 G 30 1720 en Page 37 OPMI Lumera i Device overview Fig 5 Components of the microscope 2 2 1 4 3...
Страница 39: ...Version 8 0 G 30 1720 en Page 39 OPMI Lumera i Device overview Fig 6 Controls on the microscope 2 3 4 1...
Страница 51: ...Version 8 0 G 30 1720 en Page 51 OPMI Lumera i Device overview Fig 15 Connectors on the floor stand 9 10 11 12...
Страница 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Страница 71: ...Version 8 0 G 30 1720 en Page 71 OPMI Lumera i Preparation for use Fig 24 C mount adapter 3 7 4 2 1 5 8 6...
Страница 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Страница 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Страница 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Страница 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Страница 149: ...Version 8 0 G 30 1720 en Page 149 OPMI Lumera i Device data Fig 39 Dimensional drawing of OPMI Lumera i...
Страница 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Страница 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...