7 Troubleshooting | 7.3 Column
ZEISS
5. Close the chamber door.
7.2.5 Resetting the Touch Alarm
To prevent damage, a touch alarm is integrated in the microscope. If the specimen or the speci-
men holder touches the chamber walls, the detectors, or the objective lens, the stage is stopped
immediately. An audible warning sounds and an on-screen message is displayed.
Prerequisite
ü
The EM server shows the message
WARNING Stage Touching
.
Procedure
1. To accept the warning, click
OK
.
2. Move the stage in the reverse direction away from the touch.
7.2.6 Checking the Water Flow and Temperature
Procedure
1. In the
Panel Configuration Bar
, double-click
Water Flow/Temperature
.
à
The
Water Flow/Temperature
panel is displayed.
2. Check the entries.
à
If a value is critical, it is displayed in red.
7.3 Column
7.3.1 Baking out the Gun Head
The pumping capacity of the ion getter pump decreases in the course of time, thus deteriorating
the gun vacuum. This can be remedied by an ion getter pump bakeout as a regular maintenance
procedure.
The procedure consists of the following steps:
§
§
§
7.3.1.1 Switching off the Gun
NOTICE
Schottky field emitter
If the Schottky field emitter is switched on and off too frequently or inappropriately, its lifetime
is reduced.
4
Avoid switching off the gun during the working week.
4
Use Standby mode for the weekend or a break of up to a week.
4
When using the Standby mode, activate the
Partial Vent on Standby
function.
Procedure
1. In the right part of the
Status Bar
, click
or
.
à
The pop-up menu for Vacuum, Gun and EHT activation is displayed.
2. Click
Shutdown Gun
.
3. Wait until the gun has ramped down.
à
This may take up to 5 minutes.
122
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000