ZEISS
5 Commissioning and First Operating Steps | 5.4 Acquiring an Image
Prerequisite
ü
The stage is initialized.
Procedure
1. In the GeminiSEM Control panel, select the Imaging tab.
2. In the
Detector / Active Channel
section, select
USB TV1
from the
Signal A
drop-down
list.
à
The inside of the specimen chamber is visible in the
Image Area
.
3. In the GeminiSEM Control panel, select the
Stage
tab.
4. Activate the
Track Z
checkbox.
à
The current working distance (WD) is displayed in the
Data Zone
.
5. If the
Data Zone
is deactivated, activate it via
Menu Bar > View > Data Zone > Show
Data Zone
.
6. Use the dual joystick to drive the specimen stage downwards to a low position.
NOTICE
Observe the stage movement via camera to avoid crashing.
5.4.2.3 Venting the Specimen Chamber
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4
During specimen exchange, keep the chamber door open as short as possible.
4
Do not inhale the air from within the specimen chamber.
4
Ensure that the area around the microscope is sufficiently vented.
4
If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
Procedure
1. In the GeminiSEM Control panel, select the
Vacuum
tab.
2. Click
Vent
.
à
The
Vent
message box is displayed.
3. To start venting, click
Yes
.
INFO
: If the
Stage is not initialized
system message is displayed, refer to
.
à
The specimen chamber is purged with gaseous nitrogen.
5.4.2.4 Mounting the Specimen Holder
WARNING
Suffocation hazard due to lack of oxygen
Gaseous dry nitrogen is used to vent the specimen chamber during specimen exchange. Inhal-
ing nitrogen may cause unconsciousness.
4
During specimen exchange, keep the chamber door open as short as possible.
4
Do not inhale the air from within the specimen chamber.
4
Ensure that the area around the microscope is sufficiently vented.
4
If you begin to experience symptoms of asphyxia (for example: rapid breathing, loss of
mental alertness and/or muscular coordination, depression of sensations, emotional insta-
bility, fatigue) leave the room immediately and inform the facility’s safety officer.
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000