9 Technical Data and Conformity | 9.4 Performance Data and Specifications | GeminiSEM 560
ZEISS
Air Conditioning
and Quality
Parameter
Value
Heat dissipation (ambient and water cooling)
Max. 1825 W
Typical operation: 1000 W
Temperature range for operation with indi-
cated performance (24 h per day, regardless
of whether the microscope is in operation or
switched off)
21±4 °C
Recommended best temperature stability
± 0.5 °C/h
Long-term recommended stability
2 °C/24 h
Relative humidity
< 65 %
Atmospheric pressure / altitude
≤ 2000 m above sea level to guarantee an
undisturbed operation
Pollution degree
2
Mains connection
Parameter
Value
Nominal AC voltage
208–230 VAC L1/N(L2)/PE
The provided electrical connection must be in
accordance with the applicable electrical
codes for the country of installation. In order
to avoid disturbance from other installed ma-
chines, ZEISS recommends using a separate
power connection to the main distribution
panel.
Nominal frequency
50–60 Hz
Main Power Plug
The workstation is delivered with a power
cord 3x AWG10 UL-style (5 m long) with
crimps at the wires for a fixed connection.
Those are labeled according to EU standards:
§
Black - LIVE
§
White - NEUTRAL
§
Green - PROTECTIVE EARTH
A fixed connection of the microscope re-
quires an additional mains switch in close
proximity to the microscope which complies
with IEC60947-1 and IEC60947-3 regula-
tions.
Power consumption
Max. 3600 VA, dependent on accessories
Max. current
16 A at 208–230 VAC
Addition building PE
The system must be connected to a building
earth point at all times.
Minimum cross section of an additional PE is
10 mm² Cu or 16 mm² Al
An exclusive grounding connection to earth
must be provided, i.e. the grounding terminal
must not be common to other electrical
equipment.
146