7 Troubleshooting | 7.6 Apertures
ZEISS
NOTICE
Unsuitable lubricants
When using unsuitable lubricants, the vacuum system may be contaminated.
4
Only use TEM oil 300 for lubricating.
Procedure
1. Retract the respective detector.
2. Clean the rod with isopropanol with a clean, lint-free cloth.
3. Spread some drops of TEM oil 300 across the rod.
Use a clean, lint-free cloth.
7.6 Apertures
7.6.1 Cleaning the Beamsleeve Apertures
Info
Some particles maybe visible in the SEM image on the beamsleeve aperture, especially at the
tip of the aperture. As long as these particles are not heavily charging, this will not influence
your image quality or Nano VP / XVP performance.
The procedure consists of the following steps:
§
Removing the Beamsleeve Aperture [
§
Cleaning and Re-inserting the Beamsleeve Aperture [
7.6.1.1 Removing the Beamsleeve Aperture
Procedure
1. From the
Menu Bar
, select
Tools > GoTo Panel
.
à
The
Panel Configuration Bar
is displayed showing an alphabetical list of functions.
2. Double-click
Beamsleeve Configuration
.
à
The
Beamsleeve Configuration
wizard is displayed.
3. Follow the instructions of the
Beamsleeve Configuration
wizard.
4. Carefully take out the beamsleeve aperture with the plastic tweezers from the top side.
NOTICE
If you have a BSD diode installed: Do not touch the diode surface from
bottom side, this will damage the diode.
126
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000