3 Product and Functional Description | 3.3 Optional Components and Accessories
1
Stigmator X | Stigmator Y
Shapes the beam roundness by changing the stigmation deflectors.
2
Aperture X | Aperture Y
Adjusts the mid column shift and tilt deflectors for aligning the beam along the column
axis.
3
Scan Rotate
Rotates the scanning pattern 360° continuously.
This turning knob has a push button function to deactivate the scan rotate function and
reset the scan rotation to 0°.
4
Shift X | Shift Y
Shifts the scanned region of the specimen in the X and Y directions.
5
Brightness | Contrast
§
Brightness
Adjusts the image acquisition chain offset for the currently selected detector. Each
configured detector stores its own brightness.
§
Contrast
Adjusts the gain of the currently selected detector.
6
Magnification | Reduced
§
Magnification
Adjusts the magnification of the system.
§
Reduced
Changes the scan field to a reduced area. The size of the area is determined by the
current sub scan area settings.
7
Wobble
Sweeps the acceleration voltage. If the aperture is slightly misaligned, a shift in X and/or
Y direction can be observed.
8
Freeze
Stops the scan and grabs one complete frame at the current imaging conditions.
9
Exchange | Resume
§
Exchange
Starts the pre-defined macro for specimen exchange with the airlock.
§
Resume
Starts the pre-defined macro to finish specimen exchange with the airlock.
10
Camera
Switches to chamber view.
11
Focus | Scan Speed +/−
§
Focus
Changes the focal point of the column by adjusting the magnitude of the objective
lens.
§
Scan Speed +/−
Increases (+) or decreases (−) the scan speed by doubling or halving the beam dwell
time with each click step.