5 Commissioning and First Operating Steps | 5.4 Acquiring an Image
ZEISS
5.4.4 Switching on the Gun
NOTICE
Schottky field emitter
If the Schottky field emitter is switched on and off too frequently or inappropriately, its lifetime
is reduced.
4
Avoid switching off the gun during the working week.
4
Use Standby mode for the weekend or a break of up to a week.
4
When using the Standby mode, activate the
Partial Vent on Standby
function.
Prerequisite
ü
The chamber and the gun head have been evacuated.
Procedure
1. In the right part of the
Status Bar
, verify whether the gun is switched on or off.
à
If
or
is displayed, the gun is already switched on and you can skip the fol-
lowing steps.
à
If
is displayed, the gun is switched off. Follow the operating steps within this
chapter.
2. In the GeminiSEM Control panel, select the
Vacuum
tab.
3. Verify that the
EHT Vac ready
readout is
EHT Vac ready = Yes
.
If not, the correct vacuum is not achieved. Check if the
Pump
procedure has been com-
pleted.
4. In the right part of the
Status Bar
, click
.
à
The pop-up menu for vacuum, gun, and EHT activation is displayed.
5. Click
Gun On
and monitor the gun vacuum.
à
The gun runs up.
à
This may take up to 5 minutes.
5.4.5 Switching on the EHT
When you switch on the EHT, the gun starts emitting electrons.
WARNING
Radiation hazard due to X-rays
X-rays are generated inside the microscope during operation. This is unavoidable because elec-
trons are accelerated by voltages up to 30 kV.
4
Do not remove any parts around the column and chamber that are essential for radiation
protection.
4
Use genuine ZEISS parts exclusively.
4
Ensure that all local safety and X-ray protection regulations are met.
4
Only authorized ZEISS service representatives are allowed to service the microscope.
Prerequisite
ü
The chamber and the gun head have been evacuated.
ü
The gun has been switched on.
Procedure
1. In the GeminiSEM Control panel, select the Control tab.
2. Double-click the
EHT Target
readout.
à
The
EHT Target
window is displayed.
3. In the input field, enter
10
and click
OK
.
80
Instruction Manual ZEISS GeminiSEM series | en-US | Rev. 2 | 349500-8138-000