9 Technical Data and Conformity | 9.2 Performance Data and Specifications | GeminiSEM 360
ZEISS
Info
Your ZEISS Sales & Service Partner will provide you with the detailed installation requirements.
Weight and Sizes
Main Components
Length
(mm)
Width (mm)
Height
(mm)
Weight (kg)
column
1000*
1540
1866.6
780*
Table + PC
980
980
variable
97.2
Static damper
180
180
160
12.0
Pre-vacuum pump
430
376
216
24.5
Chiller (optional, water- or
air-cooled)**
530
640
740
90
Compressor**
281
139
239
25.0
* Depending on configuration
** Actual measurements are country dependent (different manufacturers).
Air Conditioning
and Quality
Parameter
Value
Heat dissipation (ambient and water cooling)
Max. 1825 W
Typical operation: 1000 W
Temperature range for operation with indi-
cated performance (24 h per day, regardless
of whether the microscope is in operation or
switched off)
21±4 °C
Recommended best temperature stability
± 0.5 °C/h
Long-term recommended stability
2 °C/24 h
Relative humidity
< 65 %
Atmospheric pressure / altitude
≤ 2000 m above sea level to guarantee an
undisturbed operation
Pollution degree
2
Mains connection
Parameter
Value
Nominal AC voltage
208–230 VAC L1/N(L2)/PE
The provided electrical connection must be in
accordance with the applicable electrical
codes for the country of installation. In order
to avoid disturbance from other installed ma-
chines, ZEISS recommends using a separate
power connection to the main distribution
panel.
Nominal frequency
50–60 Hz
Main Power Plug
The workstation is delivered with a power
cord 3x AWG10 UL-style (5 m long) with
crimps at the wires for a fixed connection.
Those are labeled according to EU standards:
134