5 Commissioning and First Operating Steps | 5.2 Switching On the Microscope System
3. Set the
MAIN
switch to the
ON
position.
4. If the internal Emergency Off (EMO) option is
installed, press the green
S2
button to confirm
the setting of the
MAIN
switch.
5. If the external Emergency Off (EMO) option is installed, press the green
S2
START button to
reset the EMO circuit and to be able to switch on the microscope again.
5.2.2 Starting the Microscope
Info
If the system was powered off for a longer time period, the ion getter pump might fail to start.
In this case, bake-out the gun head (supervisor user rights required) or contact your local ser-
vice center. Refer to
Prerequisite
ü
The microscope is energized.
Procedure
1. At the front of the plinth, press the
ON
button.
à
The
ON
button blinks green while the system is activated.