4-12
IM DL850-03EN
Signals That Auto Setup Can Be Applied To
Frequency:
Approx. 50 Hz or higher
Absolute input voltage:
Signals whose maximum value is at least approx. 20 mV (at 1:1 setting)
Type:
Simple, repeating signals
Note
The auto setup feature may not work properly for signals that include a DC component or high-frequency
components.
Settings after the Execution of Auto Setup
Waveform Acquisition and Display Settings
T/div
The waveform with the shortest period out of the
waveforms affected by auto setup is used to set T/div.
T/div is set to the value at which 1.6 to 4 periods of the
waveform can be displayed.
START/STOP
START
Acquisition mode
Normal
Acquisition count
Infinite
Record length
10 k
Time base
Int.
Real-time HD recording
Off
Dual capturing
Off
Accumulation
Off
Vertical axis settings
V/Div
The value that causes the absolute values of the input
waveform to be between 1.6 and 4.5 div
Offset voltage
0 V
Coupling
DC
Bandwidth limit
FULL
Display on/off
Channels whose absolute input voltage values reach or
exceed 20 mV (1:1) are displayed (except for Scan).
The displays of modules that are not affected by auto
setup do not change.
DIV/Scale
DIV
Position
0.00 div
V Zoom
×1
Trigger Settings
Trigger mode
Auto
Trigger type
SIMPLE
Trigger source
The channel with the longest period out of the channels
whose amplitude is 1 div or greater
Trigger level/slope
The level between the maximum and minimum values/
rising
Hysteresis
Low
Hold-off
0.0 nsec
Trigger position
50%
Trigger delay
0.0 μsec
Computation Settings
Math
Off
The values of settings not listed here do not change.
4.5 Performing Auto Setup