6-23
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
High-Speed 10 MS/s, 12-Bit Non-Isolation Module (701255)
Specifications
Item
Specifications
Standard operating conditions
Temperature:
Humidity:
23 ±5°C
20 to 80% RH
After a 30-minute warm-up and after calibration
Effective measurement range
20 div (±10 div around 0 V, display range: 10 div, when Variable is OFF)
Number of input channels
2
Input coupling
AC, DC, and GND
Maximum sample rate
10 MS/s
Input format
Non-isolated, unbalanced
Frequency characteristics
1
(−3 dB point when sine wave of amplitude ±3 div is applied) DC to 3 MHz
Voltage-axis sensitivity setting
5 mV/div to 20 V/div (1-2-5 steps) (when using 1:1 probe attenuation)
Maximum input voltage
(at a frequency of 1 kHz or less)
Combined with the 701940(10:1):
2
600 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):
3
200 V (DC+ACpeak) (as a value that meets the safety standard)
250 V (DC+ACpeak) ( maximum allowable voltage, as a value that does not damage the
instrument when applied)
Vertical (voltage) axis accuracy
DC accuracy
1
5 mV/div to 20 V/div:
±(0.5% of 10 div)
Input connector
BNC connector (metallic type)
Input impedance
1 MΩ ± 1%, approx. 35 pF
−3 dB point when AC coupled
low frequency attenuation point
10 Hz or less (1 Hz or less when using the 701940)
Residual noise level
(Input section shorted)
±400 µV or ±0.06 div whichever is greater (Typical
4
)
A/D conversion resolution
12 bit (150 LSB/div)
Temperature coefficient
Zero point:
Gain:
5 mV/div to 20 V/div: ±(0.05% of 10 div)/°C(Typical
4
)
±(0.02% of 10 div)/°C(Typical
4
)
Bandwidth limit
Select from OFF, 500 kHz, 50 kHz, 5 kHz, and 5400 Hz
Cut-off characteristics:
−18 dB/OCT (Typical
4
)
Probe attenuation setting
Voltage probe:
1:1, 10:1, 100:1, 1000:1
Current probe:
10 A:1 V (for the 700937/701933), 100 A: 1 V (for the 701930/701931)
Compatible probes/cables
Voltage probe (10:1 passive probe): Recommended
701940, 17 to 46 pF: For measuring 600 Vpeak or less
Current probe (power can be supplied from the DL850/DL850V. Option)
700937 (15 A), 701930 (150 A), 701931 (500 A), 701933 (30 A)
High voltage differential probe (connect the GND cable provided with the probe to the DL850/
DL850V case)
700924 (1000:1, 100:1/1400 Vpeak): For measuring 1400 Vpeak or less
Connection cable (for low voltage 1:1)
366926 (non-isolated type BNC-alligator clip ×2: For measuring low voltage less than or
equal to 42 Vpeak)
1 Value measured under standard operating conditions.
H
L
2
Recommended:
Combined with the 10:1 passive probe
(701940)
H
L
701255
Input
terminal
Earth
3
Direct input
(cable that does not comply with the safety standards)
GND is connected to the case potential.
GND is connected to the case potential.
GND
GND
4 The typical value is a representative or standard value. It is not strictly warranted.
6.13 Module Specifications