App-2
IM DL850-03EN
When the Record Length Is 50 kpoint, 100 kpoint, 250 kpoint, 500 kpoint, or 1 Mpoint
Record Length
50 kpoint
100 kpoint
250 kpoint
500 kpoint
1 Mpoint
Time/div Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
Sample
Rate
(S/s)
Display
Record
Length
(Points)
100 ns
100 M
100
100 M
100
100 M
100
100 M
100
100 M
100
200 ns
100 M
200
100 M
200
100 M
200
100 M
200
100 M
200
500 ns
100 M
500
100 M
500
100 M
500
100 M
500
100 M
500
1 μs
100 M
1 k
100 M
1 k
100 M
1 k
100 M
1 k
100 M
1 k
2 μs
100 M
2 k
100 M
2 k
100 M
2 k
100 M
2 k
100 M
2 k
5 μs
100 M
5 k
100 M
5 k
100 M
5 k
100 M
5 k
100 M
5 k
10 μs
100 M
10 k
100 M
10 k
100 M
10 k
100 M
10 k
100 M
10 k
20 μs
100 M
20 k
100 M
20 k
100 M
20 k
100 M
20 k
100 M
20 k
50 μs
100 M
50 k
100 M
50 k
100 M
50 k
100 M
50 k
100 M
50 k
100 μs
50 M
50 k
100 M
100 k
100 M
100 k
100 M
100 k
100 M
100 k
200 μs
20 M
40 k
50 M
100 k
100 M
200 k
100 M
200 k
100 M
200 k
500 μs
10 M
50 k
20 M
100 k
50 M
250 k
100 M
500 k
100 M
500 k
1 ms
5 M
50 k
10 M
100 k
20 M
200 k
50 M
500 k
100 M
1 M
2 ms
2 M
40 k
5 M
100 k
10 M
200 k
20 M
400 k
50 M
1 M
5 ms
1 M
50 k
2 M
100 k
5 M
250 k
10 M
500 k
20 M
1 M
10 ms
500 k
50 k
1 M
100 k
2 M
200 k
5 M
500 k
10 M
1 M
20 ms
200 k
40 k
500 k
100 k
1 M
200 k
2 M
400 k
5 M
1 M
50 ms
100 k
50 k
200 k
100 k
500 k
250 k
1 M
500 k
2 M
1 M
100 ms
50 k
50 k
100 k
100 k
200 k
200 k
500 k
500 k
1 M
1 M
200 ms
20 k
40 k
50 k
100 k
100 k
200 k
200 k
400 k
500 k
1 M
500 ms
10 k
50 k
20 k
100 k
50 k
250 k
100 k
500 k
200 k
1 M
1 s
5 k
50 k
10 k
100 k
20 k
200 k
50 k
500 k
100 k
1 M
2 s
2 k
40 k
5 k
100 k
10 k
200 k
20 k
400 k
50 k
1 M
3 s
1 k
30 k
2 k
60 k
5 k
150 k
10 k
300 k
20 k
600 k
4 s
1 k
40 k
2 k
80 k
5 k
200 k
10 k
400 k
20 k
800 k
5 s
1 k
50 k
2 k
100 k
5 k
250 k
10 k
500 k
20 k
1 M
6 s
500
30 k
1 k
60 k
2 k
120 k
5 k
300 k
10 k
600 k
8 s
500
40 k
1 k
80 k
2 k
160 k
5 k
400 k
10 k
800 k
10 s
500
50 k
1 k
100 k
2 k
200 k
5 k
500 k
10 k
1 M
20 s
200
40 k
500
100 k
1 k
200 k
2 k
400 k
5 k
1 M
30 s
100
30 k
200
60 k
500
150 k
1 k
300 k
2 k
600 k
1 min
50
30 k
100
60 k
200
120 k
500
300 k
1 k
600 k
2 min
20
24 k
50
60 k
200
240 k
200
240 k
500
600 k
3 min
20
36 k
50
90 k
100
180 k
200
360 k
500
900 k
4 min
20
48 k
20
48 k
100
240 k
200
480 k
200
480 k
5 min
10
30 k
20
60 k
50
150 k
100
300 k
200
600 k
6 min
10
36 k
20
72 k
50
180 k
100
360 k
200
720 k
7 min
10
42 k
20
84 k
50
210 k
100
420 k
200
840 k
8 min
10
48 k
20
96 k
50
240 k
100
480 k
200
960 k
9 min
5
27 k
10
54 k
20
108 k
50
270 k
100
540 k
10 min
5
30 k
10
60 k
20
120 k
50
300 k
100
600 k
12 min
5
36 k
10
72 k
20
144 k
50
360 k
100
720 k
15 min
5
45 k
10
90 k
20
180 k
50
450 k
100
900 k
30 min
5
90 k
10
180 k
20
360 k
50
900 k
1 h
5
180 k
10
360 k
20
720 k
2 h
5
360 k
10
720 k
3 h
5
540 k
4 h
5
720 k
5 h
5
900 k
When the time axis setting is 100 ms or greater (the settings surrounded by bold lines) and the trigger mode is Auto or Auto Level,
waveforms are displayed in roll mode.
Appendix 1 Relationship between the Time Axis Setting, Record Length, and Sample Rate