6-39
IM DL850-03EN
Specifications
1
2
3
4
5
6
App
Index
Specifications by Measurement Modes
Item
Specifications
Frequency
Measurable frequency range
Selectable vertical axis
sensitivity
Minimum resolution
0.01 Hz to 200 kHz
0.1 Hz/div to 50 kHz/div (1-2-5 steps)
0.001 Hz
RPMs
Measurable RPMs range
Selectable vertical axis
sensitivity
Computing method
Selectable pulse/rotate range
0.01 rpm to 100000 rpm (where the input frequency is DC to 200 kHz).
0.1 rpm/div to 10 krpm/div (1-2-5 steps)
Computed from the frequency based on the number of pulses per rotation
RPMs = Frequency/(pulse/rotate value) × 60
1 to 99999
RPSs
Measurable RPSs range
Selectable vertical axis
sensitivity
Computing method
Selectable pulse/rotate range
0.001 rps to 2000 rps (where the input frequency is DC to 200 kHz).
0.01 rps/div to 200 rps/div (1-2-5 steps)
Computed from the frequency based on the number of pulses per rotation
RPSs = Frequency/(pulse/rotate value)
1 to 99999
Period
Measurable period range
Selectable vertical axis
sensitivity
Minimum resolution
5 µs to 50 s (where the minimum pulse width is 2 µs)
10 µs/div to 5 s/div (1-2-5 steps)
0.1 µs
Duty cycle
Measurable duty cycle range
Selectable vertical axis
sensitivity
Measurable frequency range
Measurement pulse selection
Minimum resolution
0 to 100%
1 %/div to 20 %/div (1-2-5 steps)
0.1 Hz to 200 kHz
Select positive or negative pulse
0.01%
Power supply frequency
Measurable frequency range
Selectable vertical axis
sensitivity
Center frequency setting
Minimum resolution
30 Hz to 70 Hz (when the center frequency is 50 Hz), 40 Hz to 80 Hz (when the center
frequency is 60 Hz), 380 Hz to 420 Hz (when the center frequency is 400 Hz)
0.1 Hz/div to 2 Hz/div (0.01 Hz resolution)
Select 50 Hz, 60 Hz, or 400 Hz
0.01 Hz
Pulse width
Measurable pulse width
Selectable vertical axis
sensitivity
Measurement pulse selection
Minimum resolution
2 µs to 50 s (where the input frequency is up to 200 kHz)
10 µs/div to 5 s/div (1-2-5 steps)
Select positive or negative pulse
0.1 µs
Pulse integration
Maximum pulse count
Selectable vertical axis
sensitivity
Frequency measuring range
Computation function
Selectable Unit/Pulse range
Counter reset
2×10
9
pulses
500.0E+18 value/div to 10.00E−21 value/div (1-2-5 range: total of 123 ranges)
0.1 Hz to 200 kHz (where the minimum pulse width is 2 µs)
Set the physical amount per pulse and display by converting the values intophysical values
such as distance and flow rate.
−9.9999E+30 to +9.9999E+30
Manual reset and over-limit reset
Velocity
Selectable vertical axis
sensitivity
Computing method
Selectable Distance/Pulse
range
500.0 E+18 value/div to 10.00 E-21 value/div (1-2-5 range: total of 123 ranges)
Set the amount of displacement per pulse and compute the velocity from the frequency
Automatic unit time conversion of s, min, and hour.
−9.9999E+30 to +9.9999E+30
6.13 Module Specifications