When Using a Bridge Head with a MIL Standard (MIL-C-26482) Connector Wiring
The connector on the Strain Module (701270) is a NDIS connector.
*
Use a connector adapter
cable (700935) by YOKOGAWA to make a MIL-NDIS conversion and connect to the Strain Module
(701270).
* Connector recommended by JSNDI (Japanese Society for Non-Destructive Inspection)
Connector adapter cable
(700935) 1.5 m long
Bridgehead with MIL standard
connector wiring
A
B
C
D
E
F
G
A
B
C
D
C
NDIS
MIL
B
D
A
NDIS connector pinout
MIL standard connector pinout
Corresponding
terminals
A
B
C D
E
F
A:
(positive bridge voltage)
B: Input–
(negative measurement signal)
C: Bridge–
(negative bridge voltage)
D: Input+
(positive measurement signal)
E: Floating common
F: Sense+
(positive bridge voltage sensing)
G: Sense–
(negative bridge voltage sensing)
A:
B: Signal–
C: Exc
D: Excitation–
E: NC
(no connection)
F: NC
(no connection)
Strain Module 701270
(STRAIN_NDIS)
When Using the A1002JC Connector by YOKOGAWA
You can create your own cable by using the YOKOGAWA A1002JC connector that is compatible
with the strain module and use the cable to connect a strain gauge bridge or a strain gage
transducer to the strain module.
A
A: (positive bridge voltage)
B: Input– (negative measurement signal)
C: Bridge– (negative bridge voltage)
D: Input+ (positive measurement signal)
E: Floating common
F: Sense+ (positive bridge voltage sensing)
G: Sense– (negative bridge voltage sensing)
B
C
D
E
F
G
A1002JC NDIS connector
(two in a set)
1
Pinouts as viewed in the
direction of arrow 1
Note
• The connector shell is connected to the case potential of the DL850/DL850V.
• Each of the signals from A to G is isolated within the module.
• When creating your own cable, we recommend that you use a shielded cable in order to shut out external
noise.
CAUTION
Take extra care when wiring the connectors. If the wiring is shorted or incorrect, it can damage
the DL850/DL850V or other instruments that are connected to the DL850/DL850V.
3.9 Connecting Bridgeheads
3-26
IM DL850-03EN