Maximum Input Voltages and Maximum Allowable Common Mode Voltages for the Modules
Applying a voltage exceeding the value indicated below may damage the input section. If the
frequency is above 1 kHz, damage may occur even when the voltage is below this value.
• For the 720210 (HS100M12)
Maximum input voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
1000 V (DC + ACpeak) CAT II
• With the safety cable (1:1; a combination of the 701901 and 701954)
5
200 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and 701954).
8
1000 Vrms (CAT II)
• Direct input (cable that does not comply with the safety standards)
10
42V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701250 (HS10M12)
Maximum input voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
600 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701954)
5
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
the instrument when applied)
• Direct input (cable that does not comply with the safety standards)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and 701954).
8
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
• Direct input (cable that does not comply with the safety standards)
10
42 V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701251 (HS1M16)
Maximum input voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
1
600 V (DC + ACpeak)
• Safety cable (1:1) (combined with 701954)
5
140 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
9
42V (DC + ACpeak)
Maximum allowable common mode voltage (at a frequency of 1 kHz or less)
• When used with the 700929 (10:1) isolated probe or the 701947 (100:1) isolated probe.
2
Or when used with the safety cable (1:1; a combination of the 701901 and + 701954).
8
400 Vrms (measurement category Other (O)), 300 Vrms (CAT II)
• Direct input (cable that does not comply with the safety standards)
10
42 V (DC + ACpeak) (CAT II, 30 Vrms)
• For the 701255 (NONISO_10M12)
This module is non-isolated. Be sure to fasten the module screws when measuring a
voltage above 42 V on this module. In addition, use the 701940 (10:1) dedicated non-
isolated passive probe.
Maximum input voltage (at a frequency of 1 kHz or less)
• Combined with the 701940 (10:1) passive probe
600 V (DC + ACpeak)
• Direct input (cable that does not comply with the safety standards)
9
200 V (DC + ACpeak) (as a value that meets the safety standard)
250 V (DC + ACpeak) ( maximum allowable voltage, as a value that does not damage
the instrument when applied)