8. Setting of parameters
Page 116
105204
– WTS-300-E - V4.1 - 20.01.2022
8.12 Level
“I/O test parameters” (o. parameters)
Meaning of the parameters
All inputs and outputs of the individual components can be tested in this parameter
level. For this purpose the corresponding relays are set to 'I' or '0'.
Depending on the parameterization, individual o.-parameters can be hidden.
However, the two relays for the supply of acid and alkaline detergents are locked
against each other since never acid and alkaline detergents may never enter the
tank together!
The operating level does NOT automatically jump back to the working level!
(must be exited by simultaneously pressing the two arrow keys.)
Test relay outputs
Range
Def.
Switched-on relays remain switched on until they are reset manually.
0: Off
1: On
o.1
o.2
o.3
o.4
o.5
o.6
o.7
o.8
o.9
o.10
o.11
o.12
o.15*
o.16*
o.17*
o.18*
o.19*
o.20*
Test Relay 1
Test Relay 2
Test Relay 3
Test Relay 4
Test Relay 5
Test Relay 6
Test Relay 7
Test Relay 8
Test Relay 9
Test Relay 10
Test Relay 11
Test Relay 12
Test Relay 1
– ext. I/O module
(optionally)
Test Relay 2
– ext. I/O module
(optionally)
Test Relay 3
– ext. I/O module
(optionally)
Test Relay 4
– ext. I/O module
(optionally)
Test Relay 5
– ext. I/O module
(optionally)
Test Relay 6
– ext. I/O module
(optionally)
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0 .. 1
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
* is only displayed if ext. I / O extension module is activated and parameterized.
Switch to level
“I / O test parameters”
See section 8.1
o. parameters
Code: 10E5
Working level
for normal daily operation
General cooling parameters
code: C212
Extended cooling parameters
code: D345
General cleaning parameters
code: B454
Extended cleaning parameters 1
code: C567
Extended cleaning parameters 2
code: D432
Service parameters
code: EA03
General tank monitor parameters
code: 4711
Extended tank monitor parameters
code: F94A
Alarm configuration
code: FFFF
Configuration
code: A384
I/O test parameters
code: 10E5
NOTE:
The 4 functions
- Tank monitor alarm
- system alarm
- OFF mode active
- Robot stop
maintain their state when
the I/O test parameter
level is accessed.
However, during testing,
the state of these 4 func-
tions can be changed
with
- 0: Relay de-energised
- 1: Relay energised
When leaving the level,
the original state is re-
stored.
NOTICE